In-pixel lock-in readout for wide-field nitrogen-vacancy radio-frequency spectrum analysis
Key takeaways · machine-generated
- Instantaneous bandwidth reaches 4 GHz, with frequency resolution of 7–22 MHz set by magnet position, at 5 ms readout.
- In-pixel lock-in demodulation suppresses static background and low-frequency noise before digitization, without phase synchronization to the RF carrier.
- SNR stays above 70 at -32.5 dBm input, suggesting compact room-temperature RF spectrum analysis across bands by changing only the magnet configuration.
The growing use of radio-frequency (RF) technologies requires compact, room-temperature sensors that can analyze a wide RF band in real time. Here we present a quantum RF spectrum analyzer based on the gradient-encoded electron spin resonance of a nitrogen-vacancy (NV) ensemble in diamond and read out with a 512×542-pixel lock-in camera that demodulates the signal in each pixel. On/off keying the signal under test at the camera's reference frequency lets every pixel extract the spin-resonant photoluminescence modulation. Before digitization, each pixel suppresses the static background and low-frequency noise that limit conventional camera readout. Translating a single permanent magnet tunes the encoded band continuously from 10 MHz to 32 GHz. The magnet position establishes a trade-off between instantaneous bandwidth (up to 4 GHz) and frequency resolution (7–22 MHz), with a 5 ms readout time. The wide-field readout integrates all spectral channels in parallel in a room-temperature sensor head and requires no phase synchronization to the RF carrier. The SNR remains above 70 at an applied power as low as -32.5 dBm. The magnetic-field configuration can be adapted to different RF bands without changing the pixel-parallel readout.
A wide-field quantum RF spectrum analyzer uses gradient-encoded electron spin resonance of a nitrogen-vacancy ensemble in diamond, read out by a 512×542-pixel lock-in camera. Each pixel demodulates the spin-resonant photoluminescence modulation produced by on/off keying the signal under test at the camera's reference frequency. Translating a single permanent magnet tunes the analyzed band continuously from 10 MHz to 32 GHz.